Minimum sample size: 0.5 μL. Pathlength: 1 mm (auto-ranging to 0.05 mm).
Light source: Xenon flash lamp. Detector Type: 2048-element linear silicon CCD array.
Wavelength range: 190-840 nm. Wavelength accuracy: +1 nm. Spectral Resolution: <1.8 nm (FWHM @Hg 253.7 nm).
Absorbance precision: 0.002 absorbance (1 mm path). Absorbance accuracy: ± 2% (at 0.76 absorbance at 257 nm). Absorbance range: 0.02 -300 (10 mm equivalent). Detection limit: 2 ng/μL dsDNA.
Maximum concentration: 15,000 ng/μL (dsDNA). Measurement time: < 5 seconds.
Sample pedestal. Material of construction: 303 stainless steel and quartz fiber.